OPERATION WHEN PERIPHERAL DEVICES ARE CONNECTED: Connecting to the NIDEK Auto Lensmeter (LM)
3.2.3 Operating procedure
1 After lens measurement with the LM, press the print button on the LM.
2 The TONOREF II receives lensmeter readings from the LM.
3 The LM data is printed and data is transmitted to the RT.
To RT
Interface cable
(optional)
To LM
Ferrite cores
(optional)
• When the device communicates with the LM, set the communication parameters of each
instrument as follows. See the operator’s manual for the setting method of each device.
[Settings of TONOREF II]
83. BAUD-RATE = 9600
84. BIT LENGTH = 8
86. LM DATA PRINT = YES
[Settings of NIDEK lensmeter]
PRINTER = COM PRINT
RS-232C = NIDEK
Baud Rate = 9600
Parity = Odd
Data Bits = 8
Stop Bits = 1