BEFORE USE: Measurement Screen Description
Auto button ( )
Selects the auto-tracking function and auto-shooting function.
Select auto-tracking from 3D, 2D or OFF. Select auto-shooting from ON or OFF.
Min. pupil mark
The concentric circle displayed by eight bright points indicates the minimum measurable pupil size.
If the pupil is smaller than this mark or eyelashes obscure this mark, measurement may not be possible.
Measured values
Displays the latest measured results.
Numeric values displayed to the right of “R: ” and “K: ” are the respective measurement count.
The numeric value in parentheses displayed to the right of “R: ” is a confidence index.
* “P: ” is not displayed when the 43. CONF. INDEX parameter is set to NO.
Mode mark
Indicates the set mode.
The set mode is indicated by the R/K mode mark ( ) and the NT mode mark ( ).
When the two marks are displayed at the same time, the set mode is RKT mode.
Auto-tracking mark
Indicates the setting of the auto-tracking function (alignment in the up, down, left, right, back and forth
directions and focusing in the back and forth direction).
The TONOREF II displays 3D, 2D or Manual (No indication).
Auto-shooting mark
Indicates the setting of the auto-shooting function.
CYL mode
Indicates the selected cylinder mode.
R/K button ( )
Selects a measurement mode in R/K measurement.
Select from AR/KM measurement mode, AR measurement mode or KM measurement mode. The selected
measurement mode is displayed on the screen.
Pressing the button switches the mode in the following order: AR/KM measurement mode (AR and KM continuous
measurements) AR measurement mode (AR measurement) KM measurement mode (KM measurement)
AR/KM measurement mode......
Auto-tracking in the back-and-forth, side-to-side and up-and-down directions is turned on.
Auto-tracking in the side-to-side and up-and-down directions is turned on.
(No
indication)
Manually align the device and bring the eye into focus.
Measurement starts automatically when the eye is best aligned and focused.
(No
indication)
Press the start button to start measurement.