OPERATING PROCEDURES: AR (refractive error) and KM (corneal curvature radius) Measurements
<AR measurement error messages>
Error messages are the same as those for <AR measurement error messages> of AR/KM
measurement mode.
See “2.5.1 AR (refractive error) and KM (corneal curvature radius) measurements: AR/KM
measurement mode” (page 39) for details.
2 Measurement completes.
When AI mode is YES:
When the specified number of AR measurements is performed and the data is stable (no varia-
tions), the measurement automatically completes.
*1
“<<FINISH>>” is displayed on the screen.
When AI mode is NO:
When the specified number of AR measure-
ments*
1
is performed, the measurement automati-
cally completes.
Release the start button to finish AR measure-
ment.
* If three or more measurements are obtained, the
median values are printed at the time of printing.
3 Measure the other eye in the same manner.
The subsequent procedure is the same as that for the AR/KM measurement mode.
See “2.5.1 AR (refractive error) and KM (corneal curvature radius) measurements: AR/KM
measurement mode” (page 39) for details.
*1. 3 through 10 times can be set by the 6. AR CONTINUE parameter. (Factory setting is 3 times.)