User’s and Service Guide 2-3
Getting Started
Getting Started
Figure 2-2 Connecting the Test Set to the Agilent 8753D Option 011
Network Analyzer
After all the proper connections have been made, turn on the test set
using the front panel line switch. The front panel line switch
disconnects the mains circuits from the mains supply after the EMC
filters and before other parts of the instrument.
NOTE For accurate, repeatable measurements, be sure to let the test set warm
up for at least 2 hours. It is recommended that the test set not be
turned off on a regular basis. For the most stable and accurate
measurements, leave the test set turned on at all times.
Setting the Test Set Address Switch
The test set is shipped with the GPIB (HP-IB) address set to 12, which
sets the parallel address to 0 as in Figure 2-3 on page 2-4. Refer to
“Introduction” on page 5-2 in this manual for the definition of the
parallel address.
To set the GPIB (HP-IB) address, set all five switches so that the sum of
the switches in the on or 1 position equals the desired address. In the
example below, the two switches in the on position are 8 and 4, thus the
GPIB (HP-IB) address of 12.
To set the parallel address, use only the number 1 switch. Therefore,
the only possibilities for parallel port addressing are an address of
0or1.
When GPIB (HP-IB) is used, the parallel address is ignored.
AGILENT 8753D OPTION 011
NETWORK ANALYZER
WITH
AGILENT 85047A S-PARAMETER TEST SET
AGILENT 87050A OPTION H15
PORT 1
WITH APC-7 TO
TYPE-N (F) ADAPTER
(P/N 85054-60001)
PORT 2
WITH APC-7 TO
TYPE-N (F) ADAPTER
(P/N 85054-60001)
REFLECTION TRANSMISSION