APC H15 Service Manual Page 61

  • Download
  • Add to my manuals
  • Print
  • Page
    / 86
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 60
Users and Service Guide 6-5
Servicing the Test Set
Performance Testing
Calibration
Step 1. Connect two (2) 3.5 mm flexible test port cables to Port 1 and Port 2 of
the network analyzer.
Step 2. Connect the APC-7 to 3.5 mm adapter to the end of the test Port 2 cable.
Step 3. Perform the first full two-port calibration using a 3.5 mm calibration
kit. Use the 3.5 mm standards.
Step 4. When the isolation calibration is performed, set the averaging to 16.
Step 5. Save the calibration to the internal disk and label it "Cal1".
Step 6. Perform the second full two-port calibration using a 7 mm calibration
kit. Use the 7 mm standards. The Port 1 reference plane is the 7 mm to
3.5 mm adapter.
Step 7. When the isolation calibration is performed, set the averaging to 16.
Step 8. Save the calibration to the internal disk and label it "Cal2".
Step 9. Press:
[CAL] > MORE > ADAPTER REMOVAL > HELP ADAPTER REMOVAL
Step 10. Remove the adapter and use the value recorded in step 8 of the
Adapter Delay Removal on page 6-4 for the adapter delay.
Step 11. Save the calibration and the instrument state in the internal memory
and label as "Test".
Step 12. Mark the adapter on Port 2 to indicate the port adapter reference.
Step 13. Perform an adapter removal calibration from 300 kHz to 6 GHz.
Step 14. Make sure the calibration is active.
Step 15. Save the calibration into the analyzer internal disk by pressing:
[SAVE/RECALL] > SAVE STATE
NOTE The isolation calibration routine is done with 16 averages.