APC H15 Service Manual Page 43

  • Download
  • Add to my manuals
  • Print
  • Page
    / 86
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 42
Users and Service Guide 5-9
Controlling the Test Set and Making Measurements
Commands
Calibrating the Test System
After the test set has warmed up for at least two hours, you should
calibrate it before making any measurements. Refer to your network
analyzer users guide to determine the type of calibration appropriate
for the measurements you will be making.
You will need to calibrate each measurement path separately and store
the calibration as an instrument state in the network analyzer. Refer to
your network analyzer users guide for information on how to calibrate
and store instrument states.
Refer to Figure 5-2. In this example setup the following tests will be
made:
Return loss on the DUT's input and two output ports (A and B)
Insertion loss (or gain) between the DUT's input and port A
Insertion loss (or gain) between the DUT's input and port B
Figure 5-2 Calibrating the Test System
For the best accuracy, you should perform a full two-port calibration
between test set ports 1 and 2, and again between ports 1 and 3. As
mentioned before, you need to save the calibrations as instrument
states. See your analyzer user's guide for information on calibrations
and saving instrument states.
CAUTION When performing a full two-port calibration and making subsequent
measurements, you must use the transfer switch internal to the
Agilent 8753D/E to change the direction of the RF signal path. Do not
use the test set to change the direction of the RF signal path when you
are using a full two-port calibration. Doing so will render the
calibration invalid.
AGILENT 8753E
NETWORK ANALYZER
AGILENT 87050A OPTION H15
DUT
OUT
AB
IN
CALIBRATION
PATHS
Page view 42
1 2 ... 38 39 40 41 42 43 44 45 46 47 48 ... 85 86

Comments to this Manuals

No comments