5-10 User’s and Service Guide
Controlling the Test Set and Making Measurements
Commands
Making Measurements
The following examples assume that you are using a parallel port
connection with an Agilent 8753D/E, with the test set parallel address
set to 0. Refer to “Setting the Test Set Address Switch” on page 2-3, for
information on setting the test set address.
Measuring Transmission
Refer to Figure 5-3 on page 5-11 for the following discussion. With the
Agilent 8753D/E set to measure forward transmission (S
21
), the
analyzer RF source is being output through the analyzer Port 1, and
Port 2 is set to receive the RF signal.
By using the following commands, you will connect Port 3 of the test set
to the Reflection port, and you will connect Port 7 of the test set to the
Transmission port. You will thus be measuring forward transmission
through the device under test when measuring S
21
. This will provide
you with gain or insertion loss information.
If controlling the Agilent 87050A Option H15 directly, use the following
GPIB (HP-IB) commands:
OUTPUT 716;"PARALGPIO;"
OUTPUT 716;”PARAOUT2;”
OUTPUT 716;”PARALGPIO;”
OUTPUT 716;”PARAOUT6;”
OUTPUT 712;"refl_03"
ENTER 712;”tran_07”